Hệ thống đo chất lượng không khí
Model: IQM 60
Hãng sản xuất: AEROQUAL - New Zealand
Tham khảo tại: http://www.aeroqual.com
- The IQM 60 Indoor Air Quality Monitor enables simultaneous monitoring of common indoor air quality parameters with a single instrument.
- IQM 60 Indoor Air Quality MonitorAeroqual's proprietary Analytic GSS Technology is integrated with photo-ionization detector (PID) and nondispersive infra-red (NDIR) sensors to achieve cost effective and precise measurement of gases including carbon dioxide (CO2), carbon monoxide (CO), volatile organic compounds (VOC), nitrogen dioxide (NO2) and ozone (O3). Other sensors are integrated for measuring temperature, humidity and particulate matter.
- Air quality data for each parameter is logged to a removable SD card or to a PC in real-time and can be exported for generating reports. Applications include IAQ and HVAC analysis, IAQ complaint investigation, and Sick Building Syndrome (SBS) assessment. The IQM 60 offers the complete package for performing in-depth analysis of indoor air quality.
Features:
- Multiple Gas Measurements
- Temperature and Relative Humidity
- GSS, PID and NDIR Sensor Technology
- Factory Certified – NIST Traceable
- Temperature and Humidity Sensors
- Particulate Measurement Options:
§ Particle Counter: 2-channel
§ Particulate Profiler: 8-channel
§ Particle Monitor: TSP, PM1, PM2.5, PM10
- Active-air Sampling System
- Multi-point Factory Sensor Calibration
- Zero and Span Calibration User-facility
- Real-time Data Sampling
- Removable Data Storage
- PC Data-logging Software
- Wireless Communication Options
- Compact and Portable
- Low Maintenance
Application:
- IAQ Complaint Investigation and Analysis
- HVAC System Performance Monitoring
- Sick Building Syndrome (SBS) Surveys
- Health and Comfort Assessment
- Odour Investigation and Remediation
- Testing the Efficiency of Air Purifiers
- Clean Room Monitoring and Verification
- Filtration Testing
- Residential and Commercial Buildings
- Airport Lounges and Shopping Malls
- Schools and Kindergartens
- Hospitals and Elderly Care Facilities
- Epidemiological Studies
- Re-entrainment Studies
Specifications:
Gas Modules
|
Calibrated range
|
Minimum detection limit
|
Accuracy of Factory Calibrated
|
Precision
|
Resolution
|
Carbon Dioxide CO2 (NDIR)
|
0 – 2000 ppm
|
6 ppm
|
<40 ppm + 3% of reading
|
10 ppm
|
1 ppm
|
Carbon Dioxide CO2 (NDIR)
|
0 – 5000 ppm
|
6 ppm
|
<150 ppm + 6% of reading
|
10 ppm
|
1 ppm
|
Carbon Monoxide CO (GSE)
|
0 – 100 ppm
|
0.2 ppm
|
<±2 ppm 0-20 ppm; <±10% 20-100 ppm
|
0.2 ppm
|
0.1 ppm
|
Ozone O3 (GSS)
|
0 – 0.15 ppm
|
0.001 ppm
|
<±0.005 ppm
|
0.002 ppm
|
0.001 ppm
|
Ozone O3 (GSS)
|
0 – 0.5 ppm
|
0.001 ppm
|
<±0.008 0-0.1 ppm; <±10% 0.1-0.5 ppm
|
0.005 ppm
|
0.001 ppm
|
Nitrogen Dioxide NO2 (GSS)
|
0 – 0.2 ppm
|
0.001 ppm
|
<±0.010 0-0.1 ppm; <±10% 0.1-0.2 ppm
|
0.005 ppm
|
0.001 ppm
|
Volatile Organic Compounds (PID)
(1)
|
0 – 20ppm
|
0.01 ppm
|
<±10%
|
0.02 ppm
|
0.01 ppm
|
Non-methane Hydrocarbon (GSS)
(1)
|
0 – 25ppm
|
0.1 ppm
|
<±10%
|
0.2 ppm
|
0.1 ppm
|
Volatile Organic Compounds (GSS)
(1)
|
0 – 25ppm
|
0.1 ppm
|
<±10%
|
0.2 ppm
|
0.1 ppm
|
(1) Calibrated to isobutylene (non-specific sensor)
| |||||
Temp. & Humidity Sensor
|
Range Temp.
-20°C to +100°C
|
Range RH
0-100% RH
|
Accuracy Temp. | RH @ 25°C
±0.3°C | ±2% RH
|
Temp Res.
0.01°C
|
RH Res.
0.1% RH
|
Particulate Profiler
2-channels : >0.3 µm & >2.5 µm
|
Range
0-100000 PPL
|
Particle Sizes
>0.3 & >2.5 µm
|
Accuracy
±10% to calibration aerosol
|
Flow rate
1.0 LPM
|
Resolution
1 PPL
|
Particulate Profiler
8-channels : 0.3 to 10 µm
|
Range
0-100000 PPL
|
Particle Sizes
0.3-10 µm (8)
|
Accuracy
±10% to calibration aerosol
|
Flow rate
1.0 LPM
|
Resolution
1 PPL
|
Particle Monitor
Inlet options : PM1 PM2.5 PM10 TSP
|
Range
0-2000 µg/m
|
Sensitivity
1 µg/m3
|
Accuracy
8% of NIOSH 0600
|
Precision
3 µg/m3
|
Resolution
1 µg/m3
|
Controller Module
|
Communication
RS 232
|
Data Storage
2 GB SD card
|
Data Sampling Rate
2 to 255 minutes (programmable)
|
Display
VFD 4 x 20
|
PC Software
Included
|
Gas Treatment Module
|
Sampling Pump
BLDC
|
Zero Scrubber
Built-in
|
Replaceable Zero Scrubber Media Cartridges Chemisorbant, activated carbon & Hopcalite
| ||
- Environmental Operating Range: 0 to 50°C; 10 to 90% RH (non-condensing)
- Communication Interface: RS 232 (DB9 port)/ Cable & USB adaptor supplied
- Power Supply (standard): Switch mode power supply 12V DC/ Input 90-260V AC; 47-63Hz
- Remote Communication (optional): GSM/GPRS modem; Ethernet; Wireless
- Portable Power Station (optional): Li-ion battery station & charger 222WH or other capacity
- Dimensions PM models : 378 x 236 x 132 (mm)/ Non PM models : 278 x 236 x 132 (mm)
- Enclosure Material: Aluminium & powder-coated steel
- Instrument Weight: Net 4-9 Kg (subject to configuration)
Instrument Carry Case: Impact & water resistant; 515 x 430 x 200 (mm)